Source Test Report of the Coker Steam Vent for Non-Methane and Non-Ethane Volatile Organic Compounds
Other documents for the same rulemaking (all, by stage)
References: 3 extracted, 0 linked to papers
EM SOP No.: 002-PS1 Effective Date: 15 March 2008 Expiration Date: 15 march 2009 Procedure for Low-Flow Calibration of Dry Gas Meter Using a Flow Calibrator Applicability This SOP applies to dry gas meters that are used in applications requiring a sampling rate between 1.0 and 9.0 liters per minute. This SOP is to be used on a project-specific basis when other standing calibration procedures do not adequately characterize dry gas meter performance. As this is a calibration, there is no sensitivity or detection limit associated with the SOP. A dry gas meter calibrated per this SOP is acceptable for use. A dry gas meter that does not meet the criteria of this SOP must not be used. Safety This activity will be performed within the URS facilities. Operations under this SOP must comply with URS health and safety plans. This activity has no additional safety requirements. Definitions Reserved. Equipment and Supplies Dry gas meter. DryCal Flow Calibrator –calibrated to EPA standards by third party with appropriate documentation. Stopwatch Procedure Pre sampling calibration procedure: 6.1.1 The dry gas meter must be calibrated at five different flow rates, in duplicate, by comparison of gas flow through the dry gas meter and through the DryCal flow calibrator. 6.1.1.1 6.1.2 Five different flow rates must be used. These span the range of (nominally) 1.0 to 9.0 liters per minute (0.035 to 0.32 cubic feet per minute).
USEPA. Quality Assurance Handbook for Air Pollution Measurement Systems: Volume III, Stationary Source-Specific Methods, EPA/600/r-94/038c, September 1994. 12.0 Exhibits Figure 1. Example Data Sheet
EM SOP No.: 002-PS1 Effective Date: 15 March 2008 Expiration Date: 15 march 2009 Procedure for Low-Flow Calibration of Dry Gas Meter Using a Flow Calibrator D ry Gas Meter Calibration vs D ryCal Console ID Barometric Pressure (“ Hg) DryCal ID Calibrated by Ambient Temperature (ºF) Initials Reviewed by Date Initials Date Run 1A Run 1B Run 2A Run 2B Run 3A Run 3B Run 4A Run 4B Run 5A Run 5B Flowrate (ft3/min) Flowrate (ft3/min) Flowrate (ft3/min) Flowrate (ft3/min) Flowrate (ft3/min) Flowrate (ft3/min) Flowrate (ft3/min) Flowrate (ft3/min) Flowrate (ft3/min) Flowrate (ft3/min) Nominal Flowrate (ft3/min) Start Time (hh:mm)