Approval Tests and Standards for Closed-Circuit Escape Respirators; Notice of Proposed Rulemaking
Federal Register2008-12-10 · DEPARTMENT OF HEALTH AND HUMAN SERVICES · Proposed Rule · E8-29235
Notice of proposed rulemaking.
This notice proposes updated requirements that the Department of Health and Human Service's (HHS), Centers for Disease Control and Prevention's (CDC) National Institute for Occupational Safety and Health (NIOSH) would employ to test and approve closed-circuit respirators used for escaping atmospheres considered to be immediately dangerous to life and health, including such respirators required by the Mine Safety and Health Administration (MSHA) for use in underground mines. NIOSH and MSHA jointly review and approve this type of respirator used for mine emergencies under 42 CFR pt. 84, Approval of Respiratory Protective Devices. NIOSH also approves these respirators used in other work environments where escape equipment may be provided to workers, such as vessels operated by U.S. Navy and Coast Guard personnel. The proposed rule would replace only those technical requirements in 42 CFR Part 84--Subpart H that are uniquely applicable to closed-circuit escape respirators (CCERs), a subset of the variety of escape respirators presently covered by Subpart H. All other applicable requirements of 42 CFR Part 84 would remain unchanged. The purpose of these updated requirements is to enable NIOSH and MSHA to more effectively ensure the performance, reliability, and safety of CCERs.
Original sourceXML (168 KB)PDF (239 KB)
Other documents for the same rulemaking (all, by stage)
References: 19 extracted, 7 linked to papers
Page 1Download these results: CSV · JSON
Kyriazi N, Shubilla JP [2002]. Self-contained self-rescuer field evaluation: seventh-phase results. Pittsburgh, PA: U.S. Department of Health and Human Services, Public Health Service, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 2002-127, RI 9656.
NIOSH field evaluations of certified CCERs conducted systematically and in response to the concerns of users have identified damaged respirators that failed to meet the performance criteria under which they were certified.\2\ In some instances, the designs of these respirators did not allow the user or employer to evaluate the condition of a particular respirator prior to its use in either an evacuation drill or an actual emergency.
Kyriazi N, Shubilla JP (2000). Self-contained self-rescuer field evaluation: sixth-phase results. Pittsburgh, PA: U.S. Department of Health and Human Services, Public Health Service, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 2000-128, RI 9451.
NIOSH has investigated circumstances in which units were opened and modified by unauthorized persons, effectually altering the design from the version that received NIOSH testing and certification.\11\ --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Kyriazi N (1986). Development of an automated breathing and metabolic simulator. Pittsburgh, PA: U.S. Department of the Interior, Bureau of Mines, IC 9110.
research reportNot linked to a paper · fn 12
Kyriazi N, Kovac JG, Shubilla JP, Duerr WH, Kravitz J [1986]. Self-Contained Self-Rescuer Field Evaluation: First-Year Results of 5-year Study. Pittsburgh, PA: U.S. Department of the Interior, Bureau of Mines, RI 9051.
The use of these simulators to evaluate respirator performance has been validated by NIOSH through a series of MSHA peer-reviewed studies over the past 20 years.\13\ These studies, which include side-by-side comparisons of respirator testing using three-person panels of human subjects on treadmills against testing using a breathing and metabolic simulator, demonstrate that the simulator replicates the performance of human subjects with respect to all important metabolic variables, including oxygen consumption rate, average rates of carbon dioxide production, ventilation rates, respiratory frequencies, respiratory temperatures (dry- and wet-bulb), and breathing pressures.
Kamon E, Deno S, Vercruyyen M [1984a]. Physiological responses of miners to emergency. Vol. 1—Self-contained breathing apparatus stressors. University Park, PA: The Pennsylvania State University. U.S. Bureau of Mines contract No. J0100092, p. 13.
research reportNot linked to a paper · fn 15
Fowler, B., Paul, M., Porlier, G., Elcombe, D.D., Taylor, M. 1985. A reevaluation of the minimum altitude at which hypoxic performance decrements can be detected. Ergonomics, 28(5): 781-791.
Kamon E, Deno S, Vercruyyen M [1984b]. Physiological responses of miners to emergency. Vol. 1—Self-contained breathing apparatus stressors. University Park, PA: The Pennsylvania State University. U.S. Bureau of Mines contract No. J0100092, p. 117, 119.
research reportNot linked to a paper · fn 18
Hodgson JL [1993]. Physiological costs and consequences of mine escape and rescue. University Park, PA: The Pennsylvania State University. U.S. Bureau of Mines contract No. J0345327, p. 19.
research reportNot linked to a paper · fn 19
Kamon E, Bernard T, Stein R [1975]. Steady state respiratory responses to tasks used in Federal testing of self-contained breathing apparatus. AIHA J 36:886-896.
The same man running in an upright position at 5 miles per hour is estimated to consume 2.72 liters per minute.\20\ --------------------------------------------------------------------------- ---------------------------------------------------------------------------
"Evaluation of Proposed Methods to Update Human Testing of SCBA," Turner, Beeckman, and Hodous, AIHA Journal, Volume 56, December 1995, pp 1195-1200.
journal articleNot linked to a paper · Section III, Section 84.305 (fn 23)
"Cardiorespiratory strain in jobs that require respiratory protection," Louhevaara, V., T. Tuomi, J. Smolander, O. Korhonen, et al., Int. Arch. Occup. Environ. Health. 55:195-206, 1985.
"The human energy cost of fire fighting," Lemon, P.W. and T.T. Hermiston, J. Occup. Med. 19:558-562, 1977.
journal articleNot linked to a paper · Section III, Section 84.305 (fn 23)
Vaught C, Brnich MJ, and Kellner HJ (1988). Instructional Mode and Its Effect on Initial Self-contained Self-Rescuer Donning Attempts During Training. Pittsburgh, PA: U.S. Department of the Interior, Bureau of Mines, RI 9208.
research reportNot linked to a paper · Section III, Section 84.306 (fn 25)
Zhuang Z and Bradtmiller B [2005]. Head-and-face anthropometric survey of U.S. respirator users. Journal of Occupational and Environmental Hygiene 2: 567-576.
Dayton T. Brown, Inc. Environmental Test Criteria for the Acceptability of Mine Instrumentation. USBM contract J0100040, Phase 1, Final Report DTB2GR80-0643, June 1980, 131 pp., Table 2, p. 72.
research reportNot linked to a paper · Section III, Section 84.307 (fn 26)
"The Sago Mine Disaster: A preliminary report to Governor Joe Manchin III", McAteer, J. Davitt et al., July 2006, p. 14, Buckhannon, West Virginia, http://www.wvgov.org.
MSHA Regulatory Economic Analysis, Emergency Mine Evacuation, Final Rule, December 2006 (RIN: 1219-AB46), p. 57.
government reportNot linked to a paper · Section IV.A (fn 32)
Annual Coal Report 2003. Energy Information Administration. Table 2: Coal Production and Number of Mines by State, County, and Mine Type, 2003.
datasetNot linked to a paper · Section IV.B (fn 37)
Profile of the U.S. Coal Miner 2003; http://www.nma.org/pdf/c_profile.pdf; updated October 2004.