Performance during mild acute hypoxia.
Cited in 1 place in policy documents
Approval Tests and Standards for Closed-Circuit Escape Respirators
Paul MA, Fraser WD. Performance during mild acute hypoxia. Aviation, Space, and Environmental Medicine. 1994;65(10):891-899.
As discussed above with respect to the carbon dioxide excursion limit, it allows only for brief variation to a low oxygen level, within physiologically established tolerance.\9\ To raise this excursion limit would require CCERs that would be ``overbuilt,'' resulting in unnecessarily large and/or heavy designs.