Comparison of storage phosphor computed radiography with conventional film-screen radiography in the recognition of pneumoconiosis
Open-access copyPublisher page (DOI)OpenAlex record
Cited in 1 place in policy documents
Specifications for Medical Examinations of Underground Coal Miners
Laney AS, Petsonk EL, Wolfe AL, Attfield MD [2009]. Comparison of storage phosphor computed radiography with conventional film-screen radiography in the recognition of pneumoconiosis. Eur Respir J, published ahead of print November 19, 2009.
Exact specifications for the digital imaging detector are provided because detectors must provide sufficient image size and gray scale depth to demonstrate the required subtle contrasts, and sufficient density of pixels to offer adequate resolution for the fine linear fibrotic shadows.\10\ The specification of a maximum pixel pitch of 200 [mu]m, a minimum gray-scale bit depth of 10, and spatial resolution of at least 2.5 line pairs per millimeter are based upon the existing peer-reviewed research comparing digital and traditional imaging and ensures that the use of digital radiography systems will not result in reduced ability to recognize and quantify the abnormalities.\11\ Commercially-available imaging systems are able to meet these specifications.\12\ --------------------------------------------------------------------------- ---------------------------------------------------------------------------