Policy Replication

Additional Ambient Aerosol CNC Quantitative Fit Testing Protocols: Respiratory Protection Standard

Federal Register2019-09-26 · DEPARTMENT OF LABOR; Occupational Safety and Health Administration · Rule · 2019-20686
Final rule.

OSHA is approving two additional quantitative fit testing protocols for inclusion in appendix A of the Respiratory Protection Standard. These protocols are: The modified ambient aerosol condensation nuclei counter (CNC) quantitative fit testing protocol for full-facepiece and half-mask elastomeric respirators and the modified ambient aerosol CNC quantitative fit testing protocol for filtering facepiece respirators. The protocols apply to employers in general industry, shipyard employment, and the construction industry. Both protocols are abbreviated variations of the original OSHA-approved ambient aerosol CNC quantitative fit testing protocol (often referred to as the PortaCount[supreg] protocol), but differ from the test by the exercise sets, exercise duration, and sampling sequence. These protocols will serve as alternatives to the four existing quantitative fit testing protocols already listed in appendix A of the Respiratory Protection Standard and will maintain safety and health protections for workers while providing additional flexibility and reducing compliance burdens.

RIN 1218-AC94 · Docket OSHA-2015-0015

Other documents for the same rulemaking (all, by stage)

2016-06-14 Docket documentTSI. (2014b). TSI White Paper: Analysis of the talking exercise used for respirator fit testing, July 10, 2014.
2016-10-07 Federal RegisterAdditional PortaCount® Quantitative Fit-Testing Protocols: Amendment to Respiratory Protection Standard
2016-10-07 Docket document[NIOSH] National Institute for Occupational Safety and Health. (2001). Respirator Usage in Private Sector Firms, 2001. Bureau of Labor Statistics, National Institute for Occupational Safety and Health, September 2001.
2016-10-07 Docket document[BLS] Bureau of Labor Statistics. (2014). Employer Costs for Employee Compensation, December 2014. http://www.bls.gov/news.release/archives/ecec_03112015 [See Table A].
2018-10-10 Docket document[NIOSH] National Institute for Occupational Safety and Health. (September 2003). Respirator Usage in Private Sector Firms, 2001. Bureau of Labor Statistics, National Institute for Occupational Safety and Health.
2018-11-13 Docket documentCensus Bureau (2017). 2015 County Business Patterns.

References: 7 extracted, 3 linked to papers

Page 1Download these results: CSV · JSON
Richardson, A.W. et al. (2014a), “Evaluation of a Faster Fit Testing Method for Elastomeric Half-Mask Respirators Based on the TSI PortaCount,” Journal of the International Society for Respiratory Protection 31(1): 9-22 (OSHA-2015-0015-0004)
TSI's application included three peer-reviewed articles (``the Richardson studies'') describing the accuracy and reliability of TSI's proposed protocols.\1\ The application letter also included a copy of the 2010 ANSI/AIHA (American National Standards Institute/American Industrial Hygiene Association) Z88.10 ``Respirator Fit Testing Methods'' standard (``the ANSI standard''), which contains ``Annex A2: Criteria for Evaluating New Fit Test Methods'' (``the ANSI annex'') (OSHA-2015-0015-0007).
Richardson, A.W. et al. (2013), “Evaluation of a Faster Fit Testing Method for Full-Facepiece Respirators Based on the TSI PortaCount,” Journal of the International Society for Respiratory Protection 30(2): 116-128 (OSHA-2015-0015-0005)
TSI's application included three peer-reviewed articles (``the Richardson studies'') describing the accuracy and reliability of TSI's proposed protocols.\1\ The application letter also included a copy of the 2010 ANSI/AIHA (American National Standards Institute/American Industrial Hygiene Association) Z88.10 ``Respirator Fit Testing Methods'' standard (``the ANSI standard''), which contains ``Annex A2: Criteria for Evaluating New Fit Test Methods'' (``the ANSI annex'') (OSHA-2015-0015-0007).
Richardson, A.W. et al. (2014b), “Evaluation of a Faster Fit Testing Method for Filtering Facepiece Respirators Based on the TSI PortaCount,” Journal of the International Society for Respiratory Protection 31(1): 43-56 (OSHA-2015-0015-0006)
TSI's application included three peer-reviewed articles (``the Richardson studies'') describing the accuracy and reliability of TSI's proposed protocols.\1\ The application letter also included a copy of the 2010 ANSI/AIHA (American National Standards Institute/American Industrial Hygiene Association) Z88.10 ``Respirator Fit Testing Methods'' standard (``the ANSI standard''), which contains ``Annex A2: Criteria for Evaluating New Fit Test Methods'' (``the ANSI annex'') (OSHA-2015-0015-0007).
Zhuang et al. (OSHA-2015-0015-0011)
journal articleNot linked to a paper · II.B.1.a Study Methods, fn 5
U.S. Environmental Protection Agency, “Wage Rates for Economic Analyses of the Toxics Release Inventory Program,” June 10, 2002.
government reportNot linked to a paper · fn 19
Heiden Associates, Final Report: A Study of Industry Compliance Costs Under the Final Comprehensive Assessment Information Rule, Prepared for the Chemical Manufacturers Association, December 14, 1989.
research reportNot linked to a paper · fn 19
ANSI/AIHA Z88.10-2010, Respirator Fit Testing Methods
otherNot linked to a paper · II.C Consensus Standards