Medicare Program; End-Stage Renal Disease Prospective Payment System, Payment for Renal Dialysis Services Furnished to Individuals With Acute Kidney Injury, and End-Stage Renal Disease Quality Incentive Program
This final rule updates and makes revisions to the End-Stage Renal Disease (ESRD) Prospective Payment System (PPS) for calendar year (CY) 2021. This rule also updates the payment rate for renal dialysis services furnished by an ESRD facility to individuals with acute kidney injury (AKI). In addition, this rule updates requirements for the ESRD Quality Incentive Program (QIP).
Original sourceXML (850 KB)PDF (744 KB)
Other documents for the same rulemaking
References: 75 extracted, 56 linked to papers
Medicare Improvements for Patients and Providers Act of 2008 (MIPPA)
Boschetti-de-Fierro, A., et al., "MCO Membranes: Enhanced Selectivity in High-Flux Cases," Scientific Reports, 5:18448, DOI 10.1038/srep18448
(2) CMS Analysis (a) Summary of Submitted Evidence of the Theranova Dialyzer by CMS CMS evaluated the claims and assertions made by Baxter with regard to the articles submitted by them for the Theranova Dialyzer. Patients with ESRD requiring dialysis are at high risk of mortality due to the presence of uremic toxins.\77\ However, identifying the putative uremic toxin (or toxins) has proven challenging; the European Uremic Toxin Work Group previously identified at least 90 compounds that are retained in patients undergoing dialysis.\78\ Current HD technology relies on diffusion of toxins across a semi-permeable membrane to allow for the removal of small-sized (500 Da).79 80 81 The accumulation of uremic toxins with higher molecular weight is associated with immunodeficiency, inflammation, protein-wasting, and cardiovascular complications.
Vanholder R, et al., European Uremic Toxin Work Group (EUTox). Review on uremic toxins: Classification, concentration, and interindividual variability. Kidney Int, 2003;63(5):1934-43.
(2) CMS Analysis (a) Summary of Submitted Evidence of the Theranova Dialyzer by CMS CMS evaluated the claims and assertions made by Baxter with regard to the articles submitted by them for the Theranova Dialyzer. Patients with ESRD requiring dialysis are at high risk of mortality due to the presence of uremic toxins.\77\ However, identifying the putative uremic toxin (or toxins) has proven challenging; the European Uremic Toxin Work Group previously identified at least 90 compounds that are retained in patients undergoing dialysis.\78\ Current HD technology relies on diffusion of toxins across a semi-permeable membrane to allow for the removal of small-sized (500 Da).79 80 81 The accumulation of uremic toxins with higher molecular weight is associated with immunodeficiency, inflammation, protein-wasting, and cardiovascular complications.
Macías N., et al., "Middle molecule elimination in expanded haemodialysis: only convective transport" Clin Kidney J., 2018;12(3):447-455.
García-Prieto, A., et al., "Evaluation of the efficacy of a medium cut-off dialyser and comparison with other high-flux dialysers in conventional haemodialysis and online haemodiafiltration." Clin Kidney J., 2018;11(5):742-746.
Dobre, M., et al., "Searching for Uremic Toxins" Clinical Journal of American Society of Nephrology, 2013;8(2):322-327.
Belmouaz, M., et al., "Comparison of hemodialysis with medium cut-off dialyzer and on-line hemodiafiltration on the removal of small and middle-sized molecules," Clin Nephrol, 2018;89(1):50-56.
Cordeiro, I., et al., "High-Flux versus High-Retention-Onset Membranes: In vivo Small and Middle Molecules Kinetics in Convective Dialysis Modalities," Blood Purification, 2019:1-8.
For instance, solutes such as Beta-2 microglobulin (11.8 kDa) 82 83 are associated with increased mortality.\84\ Protein-bound solutes such as indoxyl sulfate and p-cresol sulfate also appear to be poorly dialyzable and are associated with the uremic syndrome and cardiovascular disease.\85\ --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Vanholder, R., et al., "Protein-bound uremic solutes: The forgotten toxin," Kidney International, 2001.
For instance, solutes such as Beta-2 microglobulin (11.8 kDa) 82 83 are associated with increased mortality.\84\ Protein-bound solutes such as indoxyl sulfate and p-cresol sulfate also appear to be poorly dialyzable and are associated with the uremic syndrome and cardiovascular disease.\85\ --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Tanaka H, Sirich TL, Plummer NS, Weaver DS, Meyer TW. An Enlarged Profile of Uremic Solutes. PLoS One. 2015;10(8):e0135657.
Sirich, T.L, et al., "Limited reduction in uremic solute concentrations with increased dialysis frequency and time in the Frequent Hemodialysis Network Daily Trial." Kidney Int, May 2017, 91(5):1186-1192.
Clark, W.R., et al. "Uremic Toxins and their Relation to Dialysis Efficacy." Blood Purif., 2019;48(4):299-314.
Accumulation of middle molecule and protein-bound toxins may contribute to adverse outcomes among patients receiving dialysis \88\ and may explain why even a small amount of ``residual'' kidney function is strongly associated with increased survival 89 90 and higher quality of life.91 92 ---------------------------------------------------------------------------
Obi, Y., et al., "Residual Kidney Function Decline and Mortality in Incident Hemodialysis Patients," J Am Soc Nephrol., 2016;27(12):3758-3768.
Wang, A.Y. and Lai, K.N. "The importance of residual renal function in dialysis patients." Kidney Int., 2006;69(10):1726-32.
Bargman, J.M., et al., "Relative contribution of residual renal function and peritoneal clearance to adequacy of dialysis: A reanalysis of the CANUSA Study," J Am Soc Nephrol., 2001;12(10):2158-62.
Zweigart, C., et al., "Medium cut-off membranes—closer to the natural kidney removal function," Int J Artif Organs, 2017;40(7):328-334. DOI: 10.5301/uijao.5000603.
Substantial fluid losses with convection, must be replaced via infusion of typically ultrapure water and dialysis fluids.\93\ This newer technology was later supplemented by online HDF, which enables dialysis providers with ultrapure water systems to generate replacement fluid solution.
Grooteman, M.P., et al.; CONTRAST Investigators. Effect of online hemodiafiltration on all-cause mortality and cardiovascular outcomes, J Am Soc Nephrol., 2012;23(6):1087-1096.
Maduell, F., et al., ESHOL Study Group. High-efficiency postdilution online hemodiafiltration reduces all-cause mortality in hemodialysis patients, J Am Soc Nephrol., 2013;24(3):487-497.
Morena, M., et al., FRENCHIE Study Investigators. Treatment tolerance and patient-reported outcomes favor online hemodiafiltration compared to high-flux hemodialysis in the elderly, Kidney Int., 2017;91(6):1495-1509.
Ok, E., et al., Online Haemodiafiltration Study. Mortality and cardiovascular events in online haemodiafiltration (OL-HDF) compared with high-flux dialysis: Results from the Turkish OL-HDF Study, Nephrol Dial Transplant, 2013;28(1):192-202.
Krause, B., et al., "Highly selective membranes for Blood purification," Gambro Dialysatoren GmbH, Presentation abstract March 26, 2015.
Amyot, S.L, et al., "Myoglobin clearance and removal during continuous venovenous hemofiltration," Intensive Care Medicine, 1999(25):1169-1172.
In rhabdomyolysis, myoglobin clearance has been demonstrated with large pore dialyzers and HDF, but clinical benefit remains largely unproven.\103\ Similarly, HDF has historically garnered much attention in sepsis due to its ability to efficiently clear inflammatory cytokines like IL-6, but numerous studies have shown no mortality benefit in sepsis with possible downsides in the form of shortened filter life.\104\ No trials have examined the potential benefit of removing larger quantities of middle molecules than is typically achieved from high-flux membranes. --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Friedrich J.O., et al., "Hemofiltration compared to hemodialysis for acute kidney injury: Systematic review and meta-analysis," Critical Care, 2012;16:R146.
In rhabdomyolysis, myoglobin clearance has been demonstrated with large pore dialyzers and HDF, but clinical benefit remains largely unproven.\103\ Similarly, HDF has historically garnered much attention in sepsis due to its ability to efficiently clear inflammatory cytokines like IL-6, but numerous studies have shown no mortality benefit in sepsis with possible downsides in the form of shortened filter life.\104\ No trials have examined the potential benefit of removing larger quantities of middle molecules than is typically achieved from high-flux membranes. --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Kalim, S., et al., "Extended Duration Nocturnal Hemodialysis and Changes in Plasma Metabolite Profiles," Clin J Am Soc Nephrol, 2018;13(3):436-444.
Caramelo C, Just S, Gil P. Anemia in Heart Failure: Pathophysiology, Pathogenesis, Treatment and Incognitae. Rev Esp Cardiol. 2007;60(8):848-860.
Florens N, Juillard L. "Expanded Haemodialysis: News from the Field," Nephrol Dial Transplant, 2018;33:iii48-iii52.
Wolley M, Jardin M, Hutchinson, C. "Exploring the Clinical Relevance of Providing Increased Removal of Large Middle Molecules," J Am Soc Nephrol 2018;13:805-813.
Kharbanda K, et al., A Randomised Study Investigating the Effect of Medium Cut-Off Haemodialysis on Markers of Vascular Health Compared with On-Line Haemodiafiltration (MoDal Study). Manchester Metropolitan University. 2019
Kirsch AH, Lyko R, Nilsson LG., et al. Performance of hemodialysis with novel medium cut-off dialyzers. Nephrol Dial Transplant 2017;32:165-172.
Weiner DE, et al., Efficacy and Safety of Expanded Hemodialysis Enabled by a Medium Cut-Off Membrane: A Randomized Control Trial. FR-PO 488, ASN 2019.
Alarcon J, et al., Real world evidence on the impact of expanded hemodialysis (HDx) therapy on Patient Reported Outcomes (PROs): CPREXH Registry (in submission).
Sanabria RM, et al., Expanded Hemodialysis and its effects on hospitalization and medication usage: An exploratory study. (in submission).
Plumb TJ, et al., Safety and efficacy of the Tablo hemodialysis system for in-center and home hemodialysis. Hemodial Int. 2020;24(1):22-28. doi:10.1111/hdi.12795.
Kraus M, et al., A comparison of center-based vs. home-based daily hemodialysis for patients with end-stage renal disease. Hemodialysis International, 11:468-477, 2007.
Kraus M, et al.\203\ is a study involving the comparator technology known as NxStage[supreg] System, which is a portable HD unit.
Wilcox, Stephen B., et al. Results of Human Factors Testing in a Novel Hemodialysis System Designed for Ease of Patient Use. Hemodialysis International, vol. 20, no. 4, 2016, pp. 643-649.
Sehasi, Rebecca et al. Factors Associated With Discontinuation of Home Hemodialysis, American Journal of Kidney Disease, 2016;67(4):629-637.
Contributing to this low penetration rate is also a high dropout rate with the incumbent home devices of 25 percent and 35 percent at 12 and 24 months, respectively.\154\ The barriers to home dialysis adoption and retention have been well studied and include: (1) Treatment burden for patients and care partner fatigue; (2) technical challenges operating HD machine; (3) space, home modifications, and supplies management; (4) patients not wanting medical equipment in the home; and (5) safety concerns.155 156 The applicant asserted that Tablo[supreg] is the first new home HD system in over 15 years, designed to address many of the above-mentioned barriers that currently result in patients resigning themselves to in-center care and/or stopping home modalities due to the associated burden of self-managed therapy.
Seshasai, R.K., et al. The home hemodialysis patient experience: A qualitative assessment of modality use and discontinuation. Hemodialysis International, 23:139-150, 2019.
Chan, Christopher T. et al. Exploring Barriers and Potential Solutions in Home Dialysis: An NKF-KDOQI Conference Outcomes Report, American Journal of Kidney Diseases, 2019;73(3):363-371.
U.S. Department of Health and Human Services, Office of the Assistant Secretary for Planning and Evaluation, Advancing American Kidney Health, July 10, 2019.
United States Renal Data System. 2019 USRDS annual data report: Epidemiology of kidney disease in the United States. NIH, NIDDK, Bethesda, MD, 2019.
Wilk, Adam S. et al., Persistent Variation in Medicare Payment Authorization for Home Hemodialysis Treatments, Health services research vol. 53,2 (2018): 649-670.
The applicant said that among home HD patients, Medicare payment for dialysis treatments was highly variable across different regions at 3.5 to 5.7 per week.\170\ In the IDE for the Tablo[supreg] Hemodialysis System, the applicant asserted that there was effectively delivered adequacy with 4 treatments per week with an average session length of 3.4 hours, resulting in an average weekly treatment duration of ~13.6 hours.
Agency for Healthcare Quality and Research, End Stage Renal Disease in the Medicare Population: Frequency and Duration of Hemodialysis and Quality of Life Assessment, Draft Technology Assessment, November 22, 2019.
The applicant stated that Tablo[supreg] Hemodialysis System's ability to deliver adequacy on fewer treatments per week may also reduce vascular access complications due to frequent cannulation.\174\ --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Urquhart-Secord, Rachel et al. Patient and Caregiver Priorities for Outcomes in Hemodialysis: An International Nominal Group Technique Study, American Journal of Kidney Diseases, Sept. 2016;68(3):444-454.
The applicant noted that patients value having a high- quality daily life, ability to live well, and feeling empowered to control their outcomes over mortality.\175\ The applicant asserted that the use of the Tablo[supreg] Hemodialysis System at home allows patients to have an improved quality of life and control over their outcomes. --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Kidney Health Initiative, Technology Roadmap for Innovative Approaches to Renal Replacement Therapy, prepared by the Nexight Group, October 2018.
Young BA, Chan C, Blagg C, et al. How to overcome barriers and establish a successful home HD program. Clin J Am Soc Nephrol. 2012;7(12):2023-32.
In 2017, only 10.8 percent of US dialysis patients received home-based therapies.\184\ Patients and caregivers cite concerns with self-cannulation, fears of needle disconnect and complications.\185\ Home dialysis use is lower than many other rich countries.\186\ --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Wilkie M. Home dialysis—an international perspective. NDT Plus. 2011;4(Suppl 3):iii4-iii6.
In 2017, only 10.8 percent of US dialysis patients received home-based therapies.\184\ Patients and caregivers cite concerns with self-cannulation, fears of needle disconnect and complications.\185\ Home dialysis use is lower than many other rich countries.\186\ --------------------------------------------------------------------------- ---------------------------------------------------------------------------
Berger A, Edelsberg J, Inglese GW, Bhattacharyya SK, Oster G. Cost comparison of peritoneal dialysis versus hemodialysis in end-stage renal disease. American Journal of Managed Care. 2009;15:509-518.
Prior analyses have noted that home-based dialysis affords greater patient flexibility, improved quality of life,\189\ increased likelihood of employment,\190\ and improved cost.\191\ However, regarding cost comparisons, it is important to note that many cost analyses of home- based dialysis include estimates from PD.
Ash SR. The Allient dialysis system. Semin Dial. 2004;17(2):164-6.
Cafazzo JA, Leonard K, Easty AC, Rossos PG, Chan CT. Patient-perceived barriers to the adoption of Nocturnal Home Hemodialysis. Clin J Am Soc Nephrol. 2009;4:784-789.
Suri RS, Larive B, Garg AX, et al. Burden on caregivers as perceived by hemodialysis patients in the frequent Hemodialysis network (FHN) trials. Nephrol Dial Transplant. 2011;26:2316-2322.