PURPOSE: The purpose of this report is to determine if all Medtronic CRHF Therapy Delivery System (TDS) internally manufactured leads, accessory and adaptor products comply with residual requirements stated in ISO 10993-7: 2008/AC:2009 (ref. section 4.0) after being sterilized in the 3MTM 100% ethylene oxide (EO) 5XLe tabletop sterilizer systems.
BACKGROUND & PRODUCT DESCRIPTION: In 2008, a new revision of ISO 10993-7 was approved and released and conflicting national standards shall be withdrawn at the latest by October 2011. The acceptance criteria for EO and ethylene chlorohydrin (ECH) residual limits were updated from what was listed in the previous issue of ISO 10993-7:1995 (see Tables 1 and 2). In addition, a new requirement, called the Tolerable Contact Limit (TCL), was implemented to determine possible EO and ECH irritation within a patient and replaces the 250 ppm requirement Page 1 of 92 MPROC_Villalba_2014_Residuals_NonCBI_Redacted - 000001
PRODUCT STERILIZATION REPORT Doc. No.: 10173449DOC Revision: 1A Date: 04 December 2014 of AAMI TIR 19 (ref. section 4.0). All legacy internally EO sterilized CRHF TDS lead, accessory and adaptor legacy products were originally qualified to the 1995 version of ISO 10993-7. Therefore, all previously qualified (to 10993-7:1995 acceptance criteria) Medtronic CRHF TDS internally EO sterilized lead, accessory and adaptor products require assessment to ensure they meet the new residual and TCL requirements listed in ISO 10993-7:2008/AC:2009. Note: For informational purposes, ethylene glycol (EG) is monitored, but is not reported. Recent risk assessment studies have demonstrated that when EO residues are controlled as required, it is unlikely that biologically significant residues of
ISO 10993-7: 1995 Title Sterilization of health care products – Ethylene Oxide Part 1: Requirement for development, validation and routine control of a sterilization process for medical devices. Sterilization of medical devices - Requirements for medical devices to be labeled “STERILE”. Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals. Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals. Internal Repository External standard Page 14 of 92 MPROC_Villalba_2014_Residuals_NonCBI_Redacted - 000014
PRODUCT STERILIZATION REPORT Identifier ISO 10993-10:2010
AAMI TIR19: 1998 Minneapolis Chemical Technologies - REQ-090420011 Minneapolis Chemical Technologies - REQ-090421010 Minneapolis Chemical Technologies – 20911006 Minneapolis Chemical Technologies – 21118012 Minneapolis Chemical Technologies – 31209013 Minneapolis Chemical Technologies – 31201020 Minneapolis Chemical Technologies – 31222001 Minneapolis Chemical Technologies – REQ090826-007 Minneapolis Chemical Technologies – REQ090826-008 JH033041FR JH032241FR Certification 1119, Issue Level 3N Certification 1119, Issue Level 3L Certification 1119, Issue Level 3K BSH110811MM Sterilization Certification Report 1125 DL991732 BH021831FR DL952821.A BSH111461PC BL0022505 Doc. No.: 10173449DOC
PRODUCT STERILIZATION REPORT Doc. No.: 10173449DOC Revision: 1A Date: 04 December 2014
ISO 10993:1995. Comparison of this test method (CSS-1001-XXXX-0036) to the new compliant test method CHEMWI0219 per ISO 10993-7:2008/AC:2009 shows that the legacy test method is worst-case, as summarized in reliability report MDT1897325. The EO TCL results in the Table 92 using the old test method passed the ISO 10993-7:2008/AC:2009 acceptance criteria. Therefore, model 4296 is compliant to ISO 10993-7:2008/AC:2009. Page 58 of 92 MPROC_Villalba_2014_Residuals_NonCBI_Redacted - 000058 Doc. No.: 10173449DOC Revision: 1A Date: 04 December 2014