Policy Replication

Medical Gas Containers and Closures; Current Good Manufacturing Practice Requirements

Federal Register2016-11-18 · DEPARTMENT OF HEALTH AND HUMAN SERVICES; Food and Drug Administration · Rule · 2016-27838
Final rule.

The Food and Drug Administration (FDA or the Agency) is amending its current good manufacturing practice (CGMP) and labeling regulations regarding medical gases. FDA is requiring that portable cryogenic medical gas containers not manufactured with permanent gas use outlet connections have gas-specific use outlet connections that cannot be readily removed or replaced except by the manufacturer. FDA is also requiring that portable cryogenic medical gas containers and high-pressure medical gas cylinders meet certain labeling, naming, and color requirements. These requirements are intended to increase the likelihood that the contents of medical gas containers are accurately identified and reduce the likelihood of the wrong gas being connected to a gas supply system or container. FDA is also revising an existing regulation that conditionally exempts certain medical gases from certain otherwise-applicable labeling requirements in order to add oxygen and nitrogen to the list of gases subject to the exemption, and to remove cyclopropane and ethylene from the list.

RIN 0910-AC53 · Docket FDA-2005-N-0343

Other documents for the same rulemaking

2006-04-26 Docket documentReferences re Proposed rule of OMB Review for Medical Case Containers and Closures; Current Good Manufacturing Practice Requirements
2016-12-15 Regulatory impact analysisRIA for Medical Gas final rule FDA-2005-N-0343

References: 4 extracted, 0 linked to papers

Page 1Download these results: CSV · JSON
CGA M-15, Standard for Appropriate and Effective Regulations for Medical Gases within 21 CFR parts 201, 2015, and 210/211 (Compressed Gas Association 2014, 1st ed), at pages 1, 14-15, 35.
otherNot linked to a paper · X. References, ref. 1
CGA C-9, Standard Color Marking of Compressed Gas Containers for Medical Use (Compressed Gas Association 2013, 5th ed).
otherNot linked to a paper · X. References, ref. 2
CGA Safety Bulletin SB-26, Cylinder Connections on Portable Liquid Cryogenic Cylinders (Compressed Gas Association 2014, 4th ed).
otherNot linked to a paper · X. References, ref. 3
Medical Gas Containers and Closures; Current Good Manufacturing Practice Requirements, Final Regulatory Impact Analysis, Final Regulatory Flexibility Analysis, and Unfunded Mandates Reform Act Analysis, Docket No. FDA-2005-N-0343, available at http://www.fda.gov/AboutFDA/ReportsManualsForms/Reports/EconomicAnalyses/default.htm.
government reportNot linked to a paper · X. References, ref. 4 · cited link