Policy Replication

Standards for Accessible Medical Diagnostic Equipment

Federal Register2017-01-09 · ARCHITECTURAL AND TRANSPORTATION BARRIERS COMPLIANCE BOARD · Rule · 2016-31186
Final rule.

The Architectural and Transportation Barriers Compliance Board (Access Board or Board) is issuing accessibility standards for medical diagnostic equipment. The standards for medical diagnostic equipment (MDE Standards) contain minimum technical criteria to ensure that medical diagnostic equipment, including but not limited to, examination tables, examination chairs, weight scales, mammography equipment, and other imaging equipment used by health care providers for diagnostic purposes are accessible to, and usable by, individuals with disabilities. The MDE Standards will allow independent entry to, use of, and exit from the equipment by individuals with disabilities to the maximum extent possible. The MDE Standards do not impose any mandatory requirements on health care providers or medical device manufacturers. However, other agencies, referred to as enforcing authorities in the MDE Standards, may issue regulations or adopt policies that require health care providers subject to their jurisdiction to acquire accessible medical diagnostic equipment that complies with the MDE Standards.

RIN 3014-AA40 · Docket ATBCB-2012-0003

Other documents for the same rulemaking

2012-02-09 Federal RegisterMedical Diagnostic Equipment Accessibility Standards
2012-03-13 Federal RegisterMedical Diagnostic Equipment Accessibility Standards
2012-07-05 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee
2012-08-31 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee
2012-10-15 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee
2012-11-13 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee
2013-01-08 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee
2013-02-14 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee
2013-03-15 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee
2013-04-23 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee
2013-05-23 Federal RegisterRail Vehicles Access Advisory Committee
2013-05-31 Federal RegisterMedical Diagnostic Equipment Accessibility Standards Advisory Committee

References: 9 extracted, 2 linked to papers

Page 1Download these results: CSV · JSON
MDE Advisory Committee Report, 77, available at https://www.access-board.gov/guidelines-and-standards/health-care/about-this-rulemaking/advisory-committee-final-report.
government reportNot linked to a paper · M302.2.3 Size · cited link
Architectural Graphic Standards for auditorium seating
bookNot linked to a paper · M302.2.3 Size
The Wheeled Mobility Anthropometry Project recommended a toe clearance that is 5 inches deep maximum at 14 inches above the floor and a knee clearance that is 12 inches deep minimum at 28 inches above the floor.
otherNot linked to a paper · fn 10
Molly Follette Story, Erin Schwier, and June Issacson Kailes, “Perspectives of Patients with Disabilities on the Accessibility of Medical Equipment: Examination Tables, Imaging Equipment, Medical Chairs, and Weight Scales,” Disability and Health Journal 2 (2009), 169-179.
While valid and important issues about accessibility, most of these concerns are outside the purview of the Access Board as they relate to issues unrelated to the equipment itself or the built environment, and therefore, have not been addressed by the MDE Standards. In the preamble to the MDE NPRM, the Access Board identified the following barriers to accessibility, as documented in the Rehabilitation Engineering Research Center on Accessible Medical Instrument National Survey,\3\ including equipment characteristics that affect patients ability to access and use medical equipment, such as: Dimensions of the equipment (e.g., height, width, length,) contact surfaces (e.g., stiffness, comfort, color contrast), supports for transferring onto and off of equipment and positioning their bodies on the equipment (e.g., handholds, armrests, side rails), controls (e.g., ease of operation), and displays and devices (e.g., legibility and understandability).
MDE Advisory Committee Report (December 6, 2013), available at https://www.access-board.gov/guidelines-and-standards/health-care/about-this-rulemaking/advisory-committee-final-report
research reportNot linked to a paper · Section II, Rulemaking History · cited link
Wheeled Mobility Anthropometry Project, final report issued in 2010, available at http://www.udeworld.com/anthropometrics.html
research reportNot linked to a paper · fn 5 · cited link
Impact of Transfer Setup on the Performance of Independent Transfers study by the VA Pittsburgh Healthcare System in collaboration with the Human Engineering Research Laboratories at the University of Pittsburgh, available at http://herl.pitt.edu/ab/
research reportTHE IMPACT OF TRANSFER SETUP ON THE PERFORMANCE OF INDEPENDENT TRANSFERS (2011) · Section IV.B.1.d, Unobstructed Transfer
ANSI/AAMI HE 75, section 16.4.7, available at http://www.aami.org/he75
otherNot linked to a paper · Section IV.B.1.c, Transfer Surface Size · cited link
ANSI/AAMI ES60601-1:2005/(R)2012, available at http://my.aami.org
otherNot linked to a paper · Section IV.E.1.a, Structural Strength · cited link