MKS-2010 - FTIR N/A - By Difference of Typical Impurities Servomex DF-550e - Electrolytic Cell/Electrochemical This analysis of the product described herein was prepared by Praxair Distribution , Inc. using instruments whose calibration is certified using Praxai r Distribution, Inc. Reference Materials which are either prepared by weights traceable to the National Institute of Standards and Technology (NIST), Measurement Canada, or by using NIST Standard Reference Materials where available. Note: All expressions for concentration (e.g. , % or ppm) are for gas phase, by volume (e.g., ppmv) unless otherwise noted. Analytical uncertanity is expressed as a Relative % unless otherwise noted.
PGVP Number: Gas Code: E03NI76E15A0295 CC505095 124 - Tooele (SAP) - UT B72019 CO2,O2,BALN Expiration Date: Reference Number: Cylinder Volume: Cylinder Pressure: Valve Outlet: Certification Date: Feb 06, 2027 153-401417362-1 152.5 CF 2015 PSIG Feb 06, 2019 Certification performed in accordance with “EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards (May 2012)” document EPA 600/R-12/531, using the assay procedures listed. Analytical Methodology does not require correction for analytical interference. This cylinder has a total analytical uncertainty as stated below with a confidence level of 95%. There are no significant impurities which affect the use of this calibration mixture. All concentrations are on a volume/volume basis unless otherwise noted. Do Not Use This Cylinder below 100 psig, i.e. 0.7 megapascals.
CALIBRATION STANDARDS Type Lot ID NTRM NTRM 13060628 98051217 Cylinder No CC413727 SG9166746BAL Concentration Uncertainty 13.359 % CARBON DIOXIDE/NITROGEN 7.015 % OXYGEN/NITROGEN 0.6% 0.7% Expiration Date May 09, 2019 Mar 07, 2019
ANALYTICAL EQUIPMENT Instrument/Make/Model Analytical Principle Last Multipoint Calibration Horiba VIA-510 SV4MEUTJ CO2 Horiba MPA-510 W603MM58 O2 CO2 NDIR (Dixon) O2 Paramagnetic (Mason) Jan 24, 2019 Jan 10, 2019
GMIS 98.64 0.893 1684b Nox = 51.6 ppm REFERENCE VALUE ONLY The calibration results published in this certificate were obtained using equipment and standards capable of producing results that are traceable to National Institute of Standards and Technology (NIST) and through NIST to the International System of Units (SI). The expanded uncertainties, if included on this certificate, use a coverage factor of k=2 to approximate the 95% confidence level of the measurement, unless otherwise noted. If uncertainties are not included on this certificate, they are available upon request. The nitrogen used as a component or balance gas as well as the oxygen used in air mixtures meets the requirements set forth in 40CFR1065.750. This calibration certificate applies only to the item described and shall not be reproduced other than in full, without written approval from the calibration facility. Calibration certificates without signatures are not valid. This calibration meets the requirements of ISO/IEC 17025-2017. Analytical Chemist Carlie Peacock Quality Manager Allison Martinez Production Laboratory: Tier 5 Labs
PGVP Number: Gas Code: Reference Number: Cylinder Volume: Cylinder Pressure: Valve Outlet: Certification Date: May 15, 2022 163-401483044-1 144.4 CF 2015 PSIG May 15, 2019 Certification performed in accordance with “EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards (May 2012)” document EPA 600/R-12/531, using the assay procedures listed. Analytical Methodology does not require correction for analytical interference. This cylinder has a total analytical uncertainty as stated below with a confidence level of 95%. There are no significant impurities which affect the use of this calibration mixture. All concentrations are on a volume/volume basis unless otherwise noted. Do Not Use This Cylinder below 100 psig, i.e. 0.7 megapascals.