Hach Method 10133 Determination of Turbidity by Laser Nephelometry Validation Study Report. January 2000. 97 pp. [W-01-13-I-D-10]
Other documents for the same rulemaking (all, by stage)
References: 1 extracted, 0 linked to papers
SAMPLE ^ Sample pumped from source * "ftrti(_-.n«tars ara (rapt law. ami al Ihe •same t/bxtidfrt Top *iew * Figure 2 references the setup used in an on going AWWARF project, which focuses on the study of level turbidity measurements. The Project Number is 0768 and is headed by Raymond D. Letterman, Ph.D., P.E. from Syracuse University. P.25 Instrument Setup and Data Collection Procedures for the Alteirnate Test Procedure Regarding Hach Method 10133 February 9, 2000 Rev. 0.2 Overview: This document is intended to be used by both the laboratory person collecting the data and trained Hach Company Technical Personnel. This document covers instrument installation, sample setup, data collection processes and report generation for this ATP regarding Hach Method 10133. The steps in this procedure are to be followed in the order that they appear in this document. 1. Instrument Installation and Setup: The instrument installation and setup will be conducted by trained Hacli Technical Personnel. The following steps must be completed for setup. Where details are not provided here, all installation recommendations will be completed using the guidance and procedures provided in each respective instrument manual. a. Mount all instrumentation onto test fixtures. This includes: 1720D, FT660,