FGD ICP/MS Standard Operating Procedure: Inductively Coupled Plasma/Mass Spectrometry for Trace Element Analysis in Flue Gas Desulfurization Wastewaters - DCN SE06364
Other documents for the same rulemaking
References: 1 extracted, 0 linked to papers
DRAFT May 2011 The synthetic FGD matrix solution listed in Section 7.2.5 is analyzed prior to initiating sample analysis. The synthetic FGD matrix solution must be ran at least once per day (immediately after calibration) while FGD wastewaters are being analyzed, and after any major changes to instrument operating conditions. 9.4.1.1 Internal standard recovery must meet method criteria (Section 9.9) and concentrations of target elements observed should be less than their reporting limits. 9.4.1.2 In some cases it may not be possible to obtain interference check solutions that are completely free of contamination. In this case, results up to 5X the reporting limit are acceptable if it is possible to demonstrate that the isotopic ratios correspond to natural abundances. For example, the natural abundances of zinc isotopes are 64Zn 48.6%; 66Zn 27.9%; 67Zn 4.1%; 68Zn 18.8%; 70Zn 0.6%. The measured ratios are not expected to be exact, especially for low abundance isotopes, and some isotope may be masked by other elements in the solution, (for example 70Zn in the above example will be masked by 70Ge) but it must be possible to explain the